常建军,袁凤琴,杨志刚,王 佳,李 照,张利俊,徐向峰,宋晓东.蒸发光散射检测器的清洗方法[J].食品安全质量检测学报,2019,10(19):6443-6448
蒸发光散射检测器的清洗方法
Cleaning method of evaporative light scattering detector
投稿时间:2019-05-01  修订日期:2019-07-02
DOI:
中文关键词:  蒸发光散射检测器  蒸发管  清洗
英文关键词:evaporative light scattering detector  evaporation tube  cleaning
基金项目:2018年国家重点研发计划重点专项(2018YFC1604202)
作者单位
常建军 内蒙古蒙牛乳业(集团)股份有限公司中心实验室 
袁凤琴 内蒙古蒙牛乳业(集团)股份有限公司中心实验室 
杨志刚 内蒙古蒙牛乳业(集团)股份有限公司中心实验室 
王 佳 内蒙古蒙牛乳业(集团)股份有限公司中心实验室 
李 照 内蒙古蒙牛乳业(集团)股份有限公司中心实验室 
张利俊 内蒙古蒙牛乳业(集团)股份有限公司中心实验室 
徐向峰 内蒙古蒙牛乳业(集团)股份有限公司中心实验室 
宋晓东 内蒙古蒙牛乳业(集团)股份有限公司中心实验室 
AuthorInstitution
CHANG Jian-Jun Inner Mongolia Mengniu Dairy Industrial Co., Ltd 
YUAN Feng-Qin Inner Mongolia Mengniu Dairy Industrial Co., Ltd 
YANG Zhi-Gang Inner Mongolia Mengniu Dairy Industrial Co., Ltd 
WANG Jia Inner Mongolia Mengniu Dairy Industrial Co., Ltd 
LI Zhao Inner Mongolia Mengniu Dairy Industrial Co., Ltd 
ZHANG LI-Jun Inner Mongolia Mengniu Dairy Industrial Co., Ltd 
XU Xiang-Feng Inner Mongolia Mengniu Dairy Industrial Co., Ltd 
SONG Xiao-Dong Inner Mongolia Mengniu Dairy Industrial Co., Ltd 
摘要点击次数: 848
全文下载次数: 918
中文摘要:
      目的 建立一种蒸发光散射检测器的清洗方法。方法 在高温氮气流的作用下, 利用高浓度的盐酸溶液去除蒸发光散射检测器蒸发管内壁的奶垢; 依次用丙酮、乙腈、甲醇、水溶液清洗之后, 高温氮气流的作用下继续烘烤; 清洗结束后, 对蒸发光散射检测器进行面积归一化。结果 乳糖在0~4 mg/mL范围内, 浓度与峰面积采用幂函数方程绘制标准曲线, 相关系数大于0.99, 检出限为190 mg/100 g; 样品的加标平均回收率为96.45%~103.16%, 相对标准偏差为3.88%~5.18%; 对清洗前后的蒸发光散射检测器进行10次重复性测试, 相对标准偏差为0.66%, 清洗后检测信号增大2倍, 仪器噪音减少4倍。结论 本方法适用于蒸发光散射检测器的彻底清洗, 可以减少更换蒸发管的次数。
英文摘要:
      Objective To establish a cleaning method for evaporative light scattering detector. Method The high-concentration hydrochloric acid solution was used to remove the scale of the inner wall of the evaporation tube of the evaporative light scattering detector under the action of a high-temperature nitrogen stream. After washing with acetone, acetonitrile, methanol, and aqueous solution in sequence, the evaporation tube of the evaporative light scattering detector was continued to be baked under the action of high-temperature nitrogen gas. After the cleaning was completed, the area of the evaporative light scattering detector was normalized. Result When lactose was in the range of 0?4 mg/m L, the concentration curve and the peak area were plotted by the power function equation. The correlation coefficient was greater than 0.99, and the limit of detection was 190 mg/100 g. The average recoveries of the sample were 96.45%?103.16%, and the relative standard deviations were 3.88%?5.18%. The evaporative light scattering detector before and after cleaning was carried out for 10 times, the relative standard deviation was 0.66%, the detection signal was increased by 2 times after cleaning, and the instrument noise was reduced by 4 times. Conclusion The method is suitable for thorough cleaning of the evaporative light scattering detector and able to reduce the replacement time of the evaporation tube.
查看全文  查看/发表评论  下载PDF阅读器