王文珺,王汝琦,刘 磊,叶 金,郑学玲,李洁云,桑华春.赭曲霉毒素A时间分辩荧光定量检测体系适用性评价[J].食品安全质量检测学报,2018,9(6):1344-1349
赭曲霉毒素A时间分辩荧光定量检测体系适用性评价
Applicability evaluation of time-resolved fluoro-immunoassay detection system for ochratoxin A
投稿时间:2018-01-08  修订日期:2018-03-05
DOI:
中文关键词:  赭曲霉毒素A  时间分辨免疫荧光层析技术  快速检测  荧光微球
英文关键词:Ochratoxin A  time-resolved fluoro-immunoassay  rapid test  fluorescent microsphere
基金项目:
作者单位
王文珺 北京智云达科技有限公司 
王汝琦 北京智云达科技股份有限公司 
刘 磊 北京智云达科技股份有限公司 
叶 金 国家粮食局科学研究院 
郑学玲 河南工业大学粮油食品学院 
李洁云 北京智云达科技股份有限公司 
桑华春 北京智云达科技股份有限公司 
AuthorInstitution
WANG Wen-Jun Beijing Zhiyunda Science and Technology Company Limited 
WANG Ru-Qi Beijing Zhiyunda Science and Technology Company Limited 
LIU Lei Beijing Zhiyunda Science and Technology Company Limited 
YE Jin Academy of State Administration of Grain 
ZHENG Xue-Ling College of Cereal and Oils Food, Henan University of Technology 
LI Jie-Yun Beijing Zhiyunda Science and Technology Company Limited 
SANG Hua-Chun Beijing Zhiyunda Science and Technology Company Limited 
摘要点击次数: 1160
全文下载次数: 991
中文摘要:
      目的 验证赭曲霉毒素A(ochratoxin A, OTA)时间分辨荧光免疫定量检测体系对谷物中赭曲霉毒素A快速检测的适用性。方法 该时间分辨荧光免疫定量检测体系包括时间分辨荧光免疫层析检测卡和时间分辩荧光定量检测仪。时间分辨荧光速测仪内置标准曲线, 直接得出待测样品中赭曲霉毒素A的含量。对待测样品进行低、中、高3个浓度添加, 每个浓度分为7份, 由不同人员检测。同时, 对实际阳性样品进行检测。 结果 对谷物做赭曲霉毒素A添加回收率实验, 回收率在98%~113%之间; 批内变异系数<15%。在实际样品的检测中, 时间分辨荧光定量检测卡得出的数值, 与质控样本标识的值没有显著性差异。结论 时间分辨荧光定量检测系统检测快速、准确, 稳定、检测设备小型化、联网可实现数据上传, 适用于大批量谷物样品的快速检测和风险评估。
英文摘要:
      Objective To verify the applicability of the method for rapid detection of ochratoxin A (OTA) in grain by OTA time-resolved fluoro-immunoassay detection system. Methods This detection system included test strips and a detector. The OTA time-resolved fluoro-immunoassay detector had standard curve, and the content could be detected directly in samples. The samples were added at spiked levels of low, middle and high concentrations, and every level was divided into seven groups and detected by different inspectors. Then, the actual positive sample was tested. Results Recoveries were between 98%-113%, and coefficients of variation were below 15%. In the detection of actual samples, the results of the time-resolved fluorescence quantitative detection card had no significant difference from the result of the quality control sample. Conclusion OTA time-resolved fluoro-immunoassay test strip is test rapidly, accurately, and the detection system is stabilization and has small checkout equipment, moreover, the data can be uploaded by networking. It is useful to test mass grain samples rapidly and grain’s safety risk assessment.